ASS19-8-02 Comparison of Rectification Columns Educational Equipment Hydrodynamics Laboratory Equipment
1.Description
The rectification columns are used for the separation of liquid phases. They operate according to the principle of distillation. Distillation is a separation process that includes the partial evaporation of a liquid phase and the condensation of the resulting gas phase. The separation process of rectification is an energy-efficient distillation process with several stages. The substance mix recommended for the operation of the experimental plant is water-ethanol.The Comparison of rectification columns experimental plant is designed for the continuous operation of one rectification column at a time. The rectification columns are a packed column with pall rings and a sieve tray column with ten trays.
You can adjust various process parameter for the investigation of the rectification columns. These include, for example, the reflux ratio and the temperature measuring point for the temperature control. The effects of the changes are determined by means of the proportion of ethanol in the products (gravimetric measurement) thus also determining the separating capacity. For the evaluation of the experiments the software can be used to determine the theoretical separation steps by means of a McCabe-Thiele Diagram and the HTU-NTU concept.
The experimental plant is equipped with a comprehensive range of functions for measurement, control and operation that are controlled by a PLC. A touch panel displays the measured values and operating states and can be used to control the plant. At the same time, the measured values can be transmitted directly to a PC via USB where they can be analysed with the software.The steam supply is realised via the laboratory supply or the optionally available electric steam generator .
2.Specifications
- continuous rectification with packed column or sieve tray column
- variable reflux ratio
- packed column with pall rings with 10 feed levels and temperature measurement
- sieve tray column with 10 trays; each tray with feed and temperature measurement
- tanks for feed, bottom and top product made of DURAN glass and stainless steel
- operation at up to 115°C and 1,5bar
- recording of all relevant variables with more than 40 sensors
- PLC with touch panel for control of the plant
- software for data acquisition via USB under Windows 8.1, 10
3.Technical data
Packed column
-inner diameter: 100mm
-packing height: 3000mm
Sieve tray column
-inner diameter: 100mm
-number of sieve trays: 10
Feed pump
-max. flow rate: 19L/h
Tank
-feed: 20L
Measuring ranges
-temperature: 31x 0-150°C
-flow rate: 1x 1.5-20L/h (feed)
-flow rate: 1x 0.3-105L/h (distillate)
-flow rate: 1x 24-720L/h (cooling water)
pressure:
-5x 0-2.5bar (abs.)
-1x 0-5bar
-1x 0-10bar
-230V, 50Hz, 1 phase
-230V, 60Hz, 1 phase, 120V, 60Hz, 1 phase
-UL/CSA optional
Dimensions and weight
-LxWxH: 2030x850x4000mm
-Weight: approx. 480kg
Accessories
-PC with Windows recommended
-software + USB cable
-1 set of accessories
-1 set of instructional material
Electric Steam Generator
- steam generation
- electric heater
- extensive safety equipment
Technical data
-Heater
-heating power: 12,6kW
Steam generator
-steam output: 13kg/h
-operating pressure: 3…7bar
-max. pressure: 7bar
-400V, 50Hz, 3 phases
-230V, 60Hz, 3 phases
-400V, 60Hz, 3 phases
-UL/CSA optional
Dimensions and weight
-LxWxH: 680x690x1990mm
-Weight: approx. 147kg